pontiac Workshop Repair Guides

Pontiac Workshop Service and Repair Manuals

Diagram Information and Instructions|Page 11407 > < Diagram Information and Instructions|Page 11405
Page 7
background image

^

If the DMM does not display voltage refer to the next step.

4. Close each connector or switch until the DMM displays voltage in order to find which circuit is shorted.

Connector Test Adapters

NOTE:  Do not insert test equipment probes into any connector or fuse block terminal. The diameter of the test probes will deform most terminals. A
deformed terminal can cause a poor connection, which can result in system failures. Always use the J 35616-A Connector Test Adapter Kit or the J
42675 Flat Wire Probe Adapter Kit in order to frontprobe terminals. Do not use paper clips or other substitutes as they can damage terminals and cause
incorrect measurements.

Digital Multimeter

NOTE:  Refer to Test Probe Notice in Service Precautions.

IMPORTANT:  Circuits which include any solid state control modules, such as the PCM, should only be tested with a 10 megohm or higher impedance
digital multimeter such as the J 39200.

The J 39200 Instruction Manual is a good source of information and should be read thoroughly upon receipt of the DMM as well as kept on hand for
future reference.

A DMM should be used instead of a test lamp in order to test for voltage in high impedance circuits. While a test lamp shows whether voltage is present,
a DMM indicates how much voltage is present.

The ohmmeter function on a DMM shows how much resistance exists between two points along a circuit. Low resistance in a circuit means good
continuity.

IMPORTANT:  Disconnect the power feed from the suspect circuit when measuring resistance with a DMM. This prevents incorrect readings. DMMs
apply such a small voltage to measure resistance that the presence of voltages can upset a resistance reading.

Diodes and solid state components in a circuit can cause a DMM to display a false reading. To find out if a component is affecting a measurement take a
reading once, then reverse the leads and take a second reading. If the readings differ the solid state component is affecting the measurement. 

Following are examples of the various methods of connecting the DMM to the circuit to be tested:

^

Backprobe both ends of the connector and either hold the leads in place while manipulating the connector or tape the leads to the harness for
continuous monitoring while you perform other operations or test driving. Refer to Probing Electrical Connectors.

^

Disconnect the harness at both ends of the suspected circuit where it connects either to a component or to other harnesses.

^

If the system that is being diagnosed has a specified pinout or breakout box, it may be used in order to simplify connecting the DMM to the circuit
or for testing multiple circuits quickly.

Test Lamp

NOTE:  Refer to Test Probe Notice in Service Precautions.

A test lamp can simply and quickly test a low impedance circuit for voltage.

The J 34142-B Test Lamp is Micro-Pack compatible and comprised of a 12 volt light bulb with an attached pair of leads.

To properly operate this tool use the following procedure.

1. Attach one lead to ground.
2. Touch the other lead to various points along the circuit where voltage should be present.
3. When the bulb illuminates, there is voltage at the point being tested.

Using Fused Jumper Wires

IMPORTANT:  A fused jumper may not protect solid state components from being damaged.

The J 36169-A fused jumper includes small clamp connectors that provide adaptation to most connectors without damage. This fused jumper wire is
supplied with a 20 A fuse which may not be suitable for some circuits. Do not use a fuse with a higher rating than the fuse that protects the circuit being
tested.

Flat Wire Repairs

NOTE:  The flat wire within the flex wiring harness is not serviceable. If an open or short exists within the flex wiring harness the complete harness
must be replaced.

Heated Oxygen Sensor (HO2S) Wiring Repairs

NOTE:  Do not solder repairs under any circumstances as this could result in the air reference being obstructed.

If the heated oxygen sensor pigtail wiring, connector, or terminal is damaged the entire oxygen sensor assembly must be replaced. Do not attempt to
repair the wiring, connector, or terminals. In order for the sensor to function properly it must have a clean air reference. This clean air reference is
obtained by way of the oxygen sensor signal and heater wires. Any attempt to repair the wires, connectors or terminals could result in the obstruction of
the air reference and degrade oxygen sensor performance.

The following guidelines should be used when servicing the heated oxygen sensor:

Diagram Information and Instructions|Page 11407 > < Diagram Information and Instructions|Page 11405