Control line to vacuum unit for exhaust gas recirculation cooler changeover function: check |
–
| Remove the hose from the middle connection on the exhaust gas recirculation cooler change-over valve -N345--arrow-. |
–
| Connect hand vacuum pump -VAS 6213- to disconnected hose and generate a vacuum of 0.6 bar. |
–
| Observe the pressure gauge of the manual vacuum pump for 30 seconds. The vacuum must not drop. |
–
| Remove the hose of the vacuum capsule of the radiator of the exhaust gas recirculation. |
–
| Connect hand vacuum pump -VAS 6213- to vacuum unit with test hose supplied and generate a vacuum of 0.6 bar. |
Note t
| You must be able to recognise the functional sequence of the vacuum capsule and the vacuum must not drop. If this was not the case, the exhaust gas recirculation cooler system must be replaced → Chapter. |
t
| When no fault of the underpressure capsule can be recognised, the vacuum line of the exhaust gas recirculation cooler change-over valve -N345- must be replaced. |
|
|
|