t
| Hand vacuum pump with accessories -V.A.G 1390/- |
t
| Adapter -V.A.G 1598/22- |
t
| Hand multimeter -V.A.G 1526C- or multimeter -V.A.G 1715- |
t
| Adapter set -V.A.G 1594C- |
–
| Start engine and run at idling speed. |
–
| Observe positioning lever on intake manifold flap (2nd person required). |
–
| The intake manifold flap must, after switching off ignition, close and after approx. 3 seconds open again. |
If change-over does not occur, the following checks must be carried out. |
–
| Check intake manifold flap change-over mechanics for freedom of movement. To do this, operate rods by hand. |
–
| Check function of vacuum actuator using hand vacuum pump with accessories -V.A.G 1390/-. |
–
| Check that vacuum lines are connected correctly. |
If no fault is found in the mechanical components: |
–
| Check variable intake manifold flap change-over valve -N239-. |
–
| Pull connector off variable intake manifold flap change-over valve -N239-. |
|
|
|