audi Workshop Repair Guides

Audi Workshop Service and Repair Manuals

Checking intake manifold change-over valve > < Testing throttle valve positioner sender and throttle valve potentiometer
Simos injection and ignition system (4-cylinder)
Testing intake manifold change-over function
Testing intake manifold change-over function

This test should only be carried out if the engine is losing power.

The operation of the intake manifold change-over valve is dependent on the engine load and engine speed.

Testing function

  • ‒ Run engine at idling speed for 2 or 3 minutes (to build up a vacuum which is required to operate the intake manifold change-over function).
 
N24-0439
  • ‒ With engine not running and ignition switched on, start final control diagnosis and select intake manifold change-over valve -N156
    =>Page 01-55.

→ The vacuum unit for the intake manifold change-over valve should pull and release alternately (until vacuum supply in integral vacuum reservoir is used up). When the engine is not running, the vacuum is sufficient for approx. 4 change-overs. Location of vacuum unit (vacuum actuator):=>Page 24-92 )

If the change-over is not taking place, carry out the following tests:

  • ‒ Test intake manifold change-over valve -N156 (perform final control diagnosis=>Page 24-90).
  • ‒ Make sure that vacuum hoses are properly connected and free of blockages and leaks.
  • ‒ Make sure that change-over mechanism can move freely (actuate linkage by hand).

Testing vacuum system

  • ‒ Run engine at idling speed for 2 or 3 minutes (to build up a vacuum which is required to operate the intake manifold change-over function).

Note:

The vacuum reservoir is part of the intake manifold and cannot be renewed separately.

  • ‒ Switch off ignition.
  • ‒ Unplug 2-pin connector from intake manifold change-over valve.
 
A24-0075
  • ‒ → Connect one of the contacts on solenoid valve to engine earth using test cable and crocodile clamp from V.A.G 1594.
  • ‒ Connect second contact on solenoid valve to positive via remote control V.A.G 1348/3A using adapter lead V.A.G 1348/3-2 and auxiliary cable.
  • ‒ Operate remote control V.A.G 1348/3A for 2 or 3 minutes. During this time the vacuum unit should remain in the pulled-in position.

If the vacuum unit returns to rest position, test the vacuum unit as follows:

  • ‒ Disconnect vacuum hoses from vacuum unit for intake manifold change-over valve.
  • ‒ Connect hand pump V.A.G 1390 to vacuum unit.
  • ‒ Operate hand pump and observe change-over action.
  • ‒ Test vacuum unit for leaks. (The vacuum unit should not return to rest position until the hand pump is vented to atmosphere.)
  • ‒ If necessary, fit a new vacuum unit.
 
A24-0075
  • ‒ If the vacuum unit does not pull, check to ensure clear passage through -N156 as follows:
  • ‒ → Connect one of the contacts on -N156 to engine earth using test cables and crocodile clamp from V.A.G 1594.
  • ‒ Connect second contact on -N156 to positive via remote control V.A.G 1348/3A using adapter lead V.A.G 1348/3-2 and auxiliary cable.
  • ‒ Operate remote control V.A.G 1348/3A.
  • ‒ Check to ensure clear passage through valve (disconnect vacuum hoses and blow through).
  • ‒ If necessary, fit a new intake manifold change-over valve.

Checking intake manifold change-over valve > < Testing throttle valve positioner sender and throttle valve potentiometer